High flow nasal cannula system: Uses, Safety, Operation, and top Manufacturers & Suppliers

Table of Contents

Introduction

High flow nasal cannula system is a noninvasive oxygen therapy medical device designed to deliver warmed, humidified gas (air/oxygen blend) at high flow rates through a wide-bore nasal cannula. Compared with conventional oxygen delivery (standard nasal cannula or simple masks), it is built to meet higher inspiratory demand while improving patient comfort through active humidification and temperature control.

For hospitals and clinics, High flow nasal cannula system matters because it sits at a critical point in the respiratory support pathway: it can support patients who need more than low-flow oxygen while avoiding, delaying, or bridging to more invasive therapies in some care pathways. It also has meaningful operational implications—especially oxygen consumption, staffing workflows, infection control, and equipment uptime.

This article provides general, non-clinical guidance for clinicians, hospital administrators, biomedical engineers, and procurement/operations teams. You will learn what High flow nasal cannula system is used for, when it may or may not be suitable, what you need to start, how basic operation typically works, how to manage safety, how to interpret device outputs, how to troubleshoot problems, how to approach cleaning and infection control, and how the global market and supply landscape commonly looks.

In practice, you may also hear different names for similar therapy concepts. Facilities often use terms such as HFNC, high-flow nasal oxygen, HFNO, or heated humidified high-flow therapy. While these are frequently used interchangeably in everyday conversation, procurement and clinical governance teams should still verify the exact device type and intended use as described in the manufacturer’s instructions for use (IFU), because capabilities (maximum flow, oxygen blending method, sensor presence, alarms, and approved consumables) vary.

High flow therapy programs also tend to be multidisciplinary by nature. Successful deployment typically requires coordination across respiratory therapy (or nursing-led oxygen therapy teams), physicians, infection prevention, biomedical engineering, supply chain, and facilities engineering (for oxygen infrastructure). The device may look simple at the bedside, but it interacts with building utilities, consumable logistics, and training systems in ways that become very visible during patient surges or when multiple beds are placed on high-flow simultaneously.

What is High flow nasal cannula system and why do we use it?

Clear definition and purpose

High flow nasal cannula system is hospital equipment that delivers a controlled mixture of air and oxygen at high flow rates via nasal prongs, typically with:

  • Active heating (to deliver gas close to body temperature ranges)
  • Active humidification (to reduce mucosal drying and improve tolerance)
  • Controlled flow (to better match patient inspiratory flow demand)
  • A set oxygen concentration (FiO₂) through a blender or equivalent mechanism (varies by manufacturer)

From a functional standpoint, it is a respiratory support clinical device intended to improve the consistency and tolerability of oxygen delivery compared with low-flow interfaces. It is not the same as CPAP/BiPAP/NIV, although high flow therapy can provide a small degree of positive airway pressure that depends on flow, cannula size, mouth position, and patient anatomy (varies by manufacturer and patient).

A practical way to think about “high flow” is that the device is designed to deliver total flow that can approach or exceed a patient’s peak inspiratory flow in many circumstances, which helps reduce unintended dilution from room air. Adult systems commonly support a wide flow range (often up to around 60 L/min, depending on model and cannula), while pediatric and neonatal configurations use different flow ranges and cannula geometries to match smaller airways and different comfort/safety needs.

It also helps to understand what the system is controlling versus what it is not. High flow systems generally control the delivered gas conditions (flow, set oxygen concentration, and temperature/humidification targets). They generally do not control pressure in the same way a CPAP or ventilator does, and they do not directly measure the patient’s ventilation (unless integrated into a larger respiratory platform with additional monitoring).

Typical core components you may see across many designs include:

  • A flow generator (for example, a turbine drawing room air, a connection to piped medical air, or another internal flow source)
  • An oxygen inlet (wall oxygen or cylinder) and a blending mechanism (integrated or external)
  • A heated humidifier and water chamber
  • A heated-wire breathing circuit to maintain temperature and reduce excessive cooling/condensation
  • A wide-bore nasal cannula sized for the patient population
  • A control interface (screen/knobs) and an alarm system

Understanding these building blocks is useful for troubleshooting: if therapy performance is poor, the issue is often traceable to one of these “links in the chain” (gas supply, blending, heating/humidification, circuit integrity, or interface fit).

Common clinical settings

High flow nasal cannula system is commonly used in settings where patients require close observation and rapid escalation capability, including:

  • Emergency departments
  • Intensive care units (adult, pediatric, neonatal—device configuration and cannula sizing vary)
  • High-dependency or step-down units
  • Post-anesthesia care units and perioperative pathways
  • General wards with appropriate monitoring and staffing models (facility-dependent)

It may also be used in intra-facility transport contexts when teams can ensure oxygen supply adequacy, power continuity, and safe equipment securement. Transport suitability varies by manufacturer and by local protocol.

From an operations perspective, the same device can behave very differently depending on the environment. For example, step-down and ward deployment often brings new considerations such as alarm audibility in open bays, staff-to-patient ratios, availability of respiratory therapy coverage after hours, and oxygen outlet availability in older infrastructure. Some hospitals address this by limiting high-flow therapy to certain “designated zones” or by defining explicit criteria for where high-flow can be initiated and who can manage adjustments.

Key benefits in patient care and workflow

From a care delivery and workflow perspective, High flow nasal cannula system is valued because it can:

  • Deliver higher total flow than standard nasal cannula, reducing entrainment of room air and improving stability of the intended oxygen concentration
  • Improve comfort and adherence compared with tighter-fitting masks for many patients, because patients can typically speak, eat, and expectorate more easily
  • Reduce dryness and improve secretion mobilization through humidification and warming
  • Potentially reduce work of breathing in some patient populations by meeting inspiratory demand and washing out nasopharyngeal dead space (clinical response varies)

For operations leaders and biomedical engineering teams, common workflow advantages include:

  • Rapid setup compared with some noninvasive ventilation pathways (depending on local training and the specific device)
  • Standardized disposable circuits and cannula kits that simplify stock management (varies by manufacturer)
  • Clear device alarms and displayed settings that support routine bedside checks

Additional practical advantages that some facilities report—especially when training is well established—include smoother patient communication (less claustrophobia than full-face masks for some people), fewer interruptions for basic care, and the ability to combine therapy with certain mobility and rehabilitation activities when clinically appropriate. These “small” workflow details can matter in high-volume units because they influence how consistently therapy is tolerated and how often staff have to pause and restart support.

Practical limitations to understand early

High flow therapy also introduces practical constraints that matter in procurement and deployment:

  • Oxygen demand can be high, which can stress bulk oxygen systems, wall outlets, regulators, and cylinder logistics.
  • Humidification requires water management, adding consumables, workflow steps, and infection control considerations.
  • Performance depends on correct assembly, cannula sizing, and intact heated circuit function.
  • Monitoring requirements remain significant, because clinical deterioration can occur even when a device appears to be functioning normally.

Other limitations that often emerge during scale-up include:

  • Power dependency: active heating and flow generation require reliable electrical power; facilities may need clear plans for outages or emergency power circuits.
  • Noise and patient tolerance: higher flows can be noisy or uncomfortable for some patients, which can affect adherence and require careful interface fitting and reassurance.
  • Condensation management burden: warm humidified gas in cool rooms can produce “rainout,” which adds bedside tasks and can contribute to nuisance alarms if not managed systematically.
  • Consumable lock-in and cost: many systems rely on proprietary circuits/cannulas/chambers; supply disruption or price changes can impact continuity if dual-sourcing is not feasible.

When should I use High flow nasal cannula system (and when should I not)?

This section is informational only. Clinical decisions must follow local protocols, clinician judgment, and manufacturer instructions for use.

Appropriate use cases (general)

High flow nasal cannula system is commonly considered when a patient needs more support than conventional oxygen interfaces can reliably provide, especially when the goals include improved oxygen delivery and patient tolerance. Typical scenarios where clinicians may consider it include:

  • Escalation from low-flow oxygen due to persistent oxygen needs
  • Hypoxemic respiratory conditions where noninvasive oxygen support is appropriate within the facility’s care pathway
  • Post-extubation support or weaning support in certain protocols (practice varies by institution)
  • Pre-oxygenation or oxygenation support during procedures in appropriately monitored settings (protocol-dependent)
  • Pediatric and neonatal respiratory support pathways using appropriately configured systems and interfaces (device-specific)

What is “appropriate” depends heavily on the patient’s condition, staffing, monitoring capability, and the facility’s escalation options.

From a process standpoint, many facilities treat high-flow therapy as a step-up option that should come with a clear “if/then” plan: what response is expected (based on local targets), what reassessment interval is required, and what escalation pathway is triggered if response is inadequate. These operational guardrails reduce the risk that high-flow becomes a “holding pattern” rather than a purposeful support strategy with defined next steps.

Situations where it may not be suitable

High flow nasal cannula system may be unsuitable, ineffective, or operationally risky in scenarios such as:

  • Situations requiring immediate airway protection or invasive ventilation where time is critical
  • Inability to tolerate the nasal interface or cooperate with therapy (for example, severe agitation without safe mitigation strategies)
  • Anatomical limitations that prevent effective nasal cannula placement (significant nasal obstruction, certain facial/nasal trauma, or post-operative restrictions)
  • Conditions where ventilation support (not just oxygenation) is the primary requirement, depending on local pathways and clinician assessment
  • Environments where adequate monitoring and timely escalation cannot be assured

Contraindications and cautions can be device-specific. Always check the manufacturer’s labeling and your facility’s protocol.

In addition to the clinical and anatomical factors above, there are also “system” reasons therapy may be a poor fit in a specific location. Examples include wards where oxygen outlets cannot supply sustained high demand, areas where noise and alarm management are difficult, or settings where the facility cannot reliably provide the required water type and change intervals for humidification. In these cases, even a clinically appropriate therapy can become operationally unsafe if the infrastructure cannot support consistent delivery.

Safety cautions and contraindications (general, non-clinical)

From a system safety perspective, common cautions include:

  • Risk of delayed escalation if patient response is not assessed frequently and objectively.
  • Oxygen infrastructure constraints (pipeline capacity, outlet flow limits, cylinder run-time) that can cause therapy interruption.
  • Thermal and humidity risks if heater/humidifier control fails or is misassembled (comfort issues, mucosal drying, condensation).
  • Skin and pressure injury risk at the nares, cheeks, and ears from cannula and headgear.
  • Infection prevention considerations, particularly in respiratory infection seasons, where facility PPE and room-ventilation policies may impact where and how therapy is delivered.

A further consideration in many facilities is aerosol and droplet risk management. High flow therapy is an open system that can increase the velocity of exhaled gas compared with low-flow devices, and local infection prevention guidance may specify room placement, PPE requirements, and whether a patient should wear a surgical mask over the cannula when feasible. Because these policies are facility- and jurisdiction-specific, it is important that clinical teams do not rely on informal assumptions and instead follow documented infection prevention protocols.

What do I need before starting?

Required setup, environment, and accessories

At a minimum, High flow nasal cannula system typically requires:

  • A compatible flow source (integrated turbine or wall air connection, depending on the model)
  • An oxygen source (wall oxygen or cylinder with appropriate regulator)
  • An air/oxygen blender or equivalent mixing mechanism (integrated or external; varies by manufacturer)
  • A heated humidifier and water chamber (often integrated into the system)
  • A heated breathing circuit/tubing set compatible with the device
  • A wide-bore nasal cannula in the correct size range (adult/pediatric/neonatal options vary)
  • Electrical power with appropriate grounding and surge protection per facility policy
  • Patient monitoring capability aligned with your local protocol (often includes continuous pulse oximetry; additional monitoring varies)

Operationally helpful accessories include:

  • Spare water chambers and compatible sterile/distilled water supply (as specified by the manufacturer)
  • Backup circuits/cannulas for rapid changeover
  • Secure mounting options (pole mounts, carts) and cable management to reduce trip hazards
  • Oxygen analyzers or test lungs for biomedical engineering checks (facility-dependent)

For many facilities, “required setup” should also be read as “required utility performance.” High-flow devices may connect directly to wall gas outlets rather than via standard ward flowmeters, and not every outlet/regulator configuration can sustain the demand at higher therapy settings. Before deploying across a unit, it is often helpful for facilities engineering and biomedical engineering to map:

  • Which bed spaces have medical air available (if the device needs it)
  • Whether outlets and quick-connects are compatible with the device hoses
  • Any unit-level limitations (older branches, pressure drops, outlet restrictions) that could show up as low supply alarms during peak simultaneous use

Oxygen consumption planning (practical note): when a system blends air (21% oxygen) with 100% oxygen to achieve a set FiO₂, the oxygen portion can be roughly estimated for planning purposes as:

  • O₂ flow ≈ Total flow × (FiO₂ − 0.21) / 0.79

For example, at a total flow of 60 L/min and FiO₂ 0.60, the oxygen demand is approximately 60 × 0.39 / 0.79 ≈ 30 L/min of oxygen. This kind of estimate helps explain why cylinder run-time can be short at high settings and why pipeline capacity becomes a system-wide constraint when multiple beds are treated simultaneously. (Exact behavior depends on the device and gas sources; always verify with the IFU and local engineering guidance.)

Training/competency expectations

Because High flow nasal cannula system is both gas-delivery and humidification equipment, competency should cover:

  • Correct assembly of the circuit, humidifier chamber, and cannula
  • Understanding of flow, FiO₂, and temperature settings (and what the device actually controls vs what it only displays)
  • Alarm recognition and appropriate first-response actions
  • Oxygen safety basics (fire risk, cylinder safety, pipeline awareness)
  • Infection control steps for disposable components and reusable surfaces
  • Documentation expectations (settings, changes, consumable changes, and incident reporting)

For biomedical engineers and clinical engineering teams, competency often includes preventive maintenance schedules, functional checks, electrical safety testing, and verification of sensors (if present). Calibration requirements vary by manufacturer.

Many organizations also build competency around workflow consistency, not just technical assembly. Examples include: standardizing where consumables are stored, defining who is responsible for water chamber changes, clarifying how therapy is handed over during transfers, and ensuring staff know what backup oxygen delivery method to apply immediately if the high-flow device alarms or fails. These “last mile” workflow steps often determine whether high-flow therapy is resilient during night shifts and high census periods.

Pre-use checks and documentation

A practical pre-use checklist (adapt to local policy and the IFU) often includes:

  • Confirm the device is within its preventive maintenance interval and has passed recent safety testing.
  • Inspect power cord, plug, and casing for damage; confirm the device boots without fault codes.
  • Verify gas supply availability and stability (oxygen and air pressures if using wall supplies).
  • Confirm the humidifier chamber is correctly seated and filled to the indicated level using the specified water type.
  • Ensure the heated circuit is correctly connected and that temperature control sensors/cables (if present) are properly engaged.
  • Check that the cannula package is intact and that the size is appropriate for the patient group.
  • Verify alarm audibility and that alarm limits (if configurable) align with facility policy.
  • Document baseline settings, start time, device ID/serial (facility-dependent), and initial patient tolerance per protocol.

Additional checks that can reduce avoidable interruptions include confirming that the water chamber and circuit are within any stated shelf-life/expiry, checking for visible cracks in the chamber (hairline fractures can leak under heat), ensuring the device air inlet/filter (if present) is not obstructed, and verifying that the device is positioned to avoid blocked vents. These checks are especially relevant for devices that draw room air via a turbine, where a blocked inlet can cause flow limitation alarms.

How do I use it correctly (basic operation)?

The exact workflow varies by manufacturer, model, and whether the system uses wall air, an internal turbine, or an external blender. The steps below describe a common, general sequence.

Basic step-by-step workflow

  1. Confirm readiness and protocol alignment
    Ensure the care area has monitoring, staffing, and escalation pathways aligned with your facility’s policy for High flow nasal cannula system use.

  2. Select the correct patient interface
    Choose a cannula size designed for the patient population (adult/pediatric/neonatal). A poor fit can reduce comfort and therapy effectiveness and may increase pressure injury risk.

  3. Position and power the device
    Place the unit on a stable cart or pole mount, route cables/tubing to reduce trip hazards, and connect to mains power (and battery/backup if applicable).

  4. Connect gas supplies
    Connect oxygen (and air, if required) using the manufacturer-specified hoses and connectors. Confirm supply pressure indicators and check for leaks. Misconnection risks exist in facilities; connector standards and labeling are important.

  5. Install humidifier chamber and circuit
    Insert the water chamber, fill to the marked level with the specified water type, and connect the heated circuit. Ensure temperature probes/heater wires (if part of the design) are properly seated.

  6. Warm-up and system self-check
    Many systems perform self-tests and may require a brief warm-up to reach temperature/humidity targets. Follow on-screen prompts and confirm there are no active faults.

  7. Set key parameters
    Typical adjustable parameters include:

  • Flow (L/min): total gas flow delivered to the patient
  • FiO₂ (fraction of inspired oxygen): set oxygen concentration (usually 0.21–1.0, depending on device configuration)
  • Temperature (°C): gas temperature target (commonly set within the low-to-mid 30s °C up to ~37 °C; varies by manufacturer and protocol)
  1. Apply cannula and start therapy
    Fit the cannula, secure headgear, confirm comfort, and ensure tubing is not pulling on the patient’s face. Observe initial tolerance and check that expected flow is being delivered.

  2. Ongoing monitoring and adjustments
    Reassess patient status and device function at intervals defined by local protocol. Document changes in settings, patient tolerance, and any alarms/events.

Operationally, many teams find it helpful to briefly explain to the patient what to expect (warm air sensation, higher flow, possible noise) and to check early for comfort issues such as nasal irritation or strap pressure. Comfort is not just a “nice to have”—poor tolerance can lead to frequent cannula removal, therapy interruptions, and inconsistent oxygenation support.

Typical settings and what they generally mean

While exact numbers are protocol- and patient-dependent, these concepts are consistent across most systems:

  • Higher flow generally increases the system’s ability to meet inspiratory demand and can improve stability of the intended FiO₂, but it also increases oxygen consumption and may reduce comfort if too high.
  • Higher FiO₂ increases delivered oxygen concentration, but it also increases oxygen supply requirements and should be managed within clinical governance frameworks.
  • Higher temperature/humidity typically improves mucosal comfort and secretion management, but it can increase condensation in cool rooms or with long circuits.

Adult systems commonly support flows up to around 60 L/min, but maximum flow varies by manufacturer and may differ by cannula type and patient population.

From a bedside workflow perspective, remember that flow, FiO₂, and temperature interact with the room environment and patient behavior. For example, a cool room can increase rainout even at moderate temperature settings, and an open mouth posture can influence how flow is experienced. These are not “faults” with the device, but they are variables that can affect tolerance and the frequency of condensation-related tasks.

Calibration and verification (if relevant)

Some High flow nasal cannula system designs include oxygen sensors or other measurement features that require periodic checks or calibration. Others control FiO₂ through fixed-ratio blending without a user-calibrated sensor. Always treat this as device-specific:

  • If the device has a sensor: follow the IFU for calibration frequency, gas requirements, and acceptance criteria.
  • If the device does not measure delivered FiO₂: verify the gas supply setup and blender function (if external), and understand that “set FiO₂” is a controlled input rather than a measured patient-delivered value.

Biomedical engineering teams should align verification methods with local quality systems and manufacturer recommendations.

Where a measured oxygen sensor is present, facilities often incorporate verification into routine preventive maintenance or commissioning checks (especially after repair or software updates). Even when sensors are stable, documenting calibration status can reduce uncertainty during audits and can help clinicians trust the displayed values—particularly in environments where external oxygen analysis is not readily available.

How do I keep the patient safe?

Patient safety with High flow nasal cannula system is a blend of correct device operation, appropriate monitoring, and disciplined escalation planning. The aim is to avoid both device-related harm (burns, drying, pressure injury, oxygen/fire risk) and process-related harm (delayed recognition of deterioration).

Safety practices and monitoring

Common safety practices include:

  • Use monitoring appropriate to the care area and patient acuity (facility policy varies).
  • Ensure baseline observations are recorded before initiation and trends are reviewed after any setting change.
  • Confirm the cannula is correctly positioned and that the patient is comfortable enough to continue therapy.
  • Reassess frequently for changes in work of breathing, mental status, secretion handling, and overall trajectory, not only oxygen saturation.

High flow therapy can improve oxygenation while underlying deterioration continues; monitoring should focus on the whole clinical picture per protocol.

From a systems perspective, high-flow safety improves when teams standardize who is responsible for reassessment and when it must occur (for example, time-based checks after initiation and after adjustments). Clear handover communication (current settings, water chamber status, recent alarms, and any skin issues) reduces the chance that small problems accumulate into a therapy interruption or an unrecognized deterioration.

Interface safety and comfort

To reduce preventable harm:

  • Select the correct cannula size; oversized prongs can increase pressure and nasal trauma, while undersized prongs may reduce effective flow delivery.
  • Protect skin at contact points (cheeks, behind ears) using facility-approved dressings if indicated.
  • Avoid excessive strap tension; secure tubing to prevent drag.
  • Manage dryness and epistaxis risk by ensuring humidification is functioning as intended and by following facility guidance for nasal care.

Where skin protection products are used, ensure they are compatible with oxygen-enriched environments and local infection prevention policy. For example, facilities often avoid petroleum-based products near oxygen delivery because of fire safety concerns, and they standardize which barrier films or dressings are allowed for respiratory devices.

Humidification and condensation management

Humidification improves tolerance but introduces new hazards:

  • Ensure the humidifier chamber is filled correctly and seated properly; an improperly seated chamber can lead to alarms, poor humidification, or leaks.
  • Watch for condensation (“rainout”) especially in cool rooms; pooled condensate can occlude flow or be inadvertently delivered toward the patient if tubing is lifted or tipped.
  • Use water traps or drainage steps only if supported by the manufacturer’s design; improvisation can introduce infection risk and flow resistance.

A practical bedside habit is to keep the circuit routed so that any condensate tends to collect away from the patient interface, and to avoid sudden lifting of the tubing that could move pooled water toward the cannula. When condensate needs to be handled, treat it as potentially contaminated and follow local policy for PPE and disposal.

Alarm handling and human factors

Alarm safety is not just technical—it is behavioral:

  • Do not silence alarms without investigating the cause and documenting actions per policy.
  • Treat repeated nuisance alarms as a quality issue: check assembly steps, staff training, consumable compatibility, and alarm configuration (if adjustable).
  • Standardize setup across units where possible to reduce variability and cognitive load.
  • Use clear labeling for gas hoses and outlets; misconnection risk is a real safety and downtime driver.

It can also be helpful for units to maintain a short “alarm quick guide” based on their specific device model (for example: low temperature, high temperature, low water, disconnection, low supply pressure). When new staff rotate into high-flow-capable areas, a model-specific guide can reduce response time and prevent incorrect fixes such as swapping cannulas when the real issue is a low gas supply.

Oxygen and fire safety (system-level)

High flow oxygen therapy increases oxygen enrichment around the patient:

  • Follow facility oxygen/fire safety rules, including ignition source control and signage where required.
  • Ensure cylinders are secured and regulators are appropriate for expected flow demand.
  • Engage facilities engineering when deploying high-flow therapy at scale; pipeline capacity and bulk oxygen systems can be limiting factors, especially during seasonal surges.

In addition, ensure staff understand that oxygen enrichment can make materials ignite more easily and burn more vigorously. Strict control of ignition sources (smoking, open flames, certain personal devices per policy) and correct cylinder handling are essential operational safeguards, not optional extras.

Emphasize protocols and manufacturer guidance

High flow nasal cannula system is a regulated medical device. Safe use depends on:

  • Manufacturer instructions for use (IFU)
  • Local clinical governance and escalation pathways
  • Competency-based training and periodic refreshers
  • Biomedical engineering preventive maintenance and post-repair verification

How do I interpret the output?

High flow nasal cannula system typically provides device status information rather than direct measurement of patient physiology. Interpreting outputs correctly means understanding what the machine controls, what it measures (if anything), and what it cannot know.

Types of outputs/readings you may see

Depending on the model, the device may display:

  • Set flow (L/min)
  • Set FiO₂ (or oxygen percentage)
  • Temperature target and/or measured outlet temperature
  • Humidification status (heat level, warm-up indicator, water level alarms)
  • Gas supply status (oxygen/air supply pressure alarms, disconnect indicators)
  • System fault codes and maintenance prompts

Some devices include an oxygen sensor and may display measured oxygen concentration within the device. Others display only the set value. This is a critical distinction and varies by manufacturer.

Some devices also display “actual” or “delivered” values (for example, delivered temperature or achieved flow) and may indicate whether the system is still in a warm-up phase. Understanding whether a displayed number is a setpoint or a measured value is important for troubleshooting and documentation, especially when staff are handing over care between units.

How clinicians typically interpret them (general)

In routine practice, teams generally:

  • Confirm the device is delivering the intended settings (flow/FiO₂/temperature) and that no alarms are active.
  • Correlate device settings with patient response using independent monitoring (pulse oximetry, respiratory rate, clinical assessment, and lab values as ordered).
  • Use trends, not single data points, to judge whether the patient appears to be stabilizing, improving, or deteriorating under the current support strategy.

Common pitfalls and limitations

Avoid these frequent interpretation errors:

  • Assuming “set FiO₂” equals “delivered FiO₂ at the alveoli”; patient factors and interface leak can affect actual inspired oxygen.
  • Confusing oxygen flow (from a wall flowmeter) with total flow delivered by the system; High flow nasal cannula system typically delivers a blended total flow that is not the same as oxygen-only flow.
  • Over-relying on improved oxygen saturation while missing signs of increased fatigue, rising work of breathing, or reduced consciousness.
  • Ignoring room temperature effects on condensation and humidification performance, which can influence comfort and circuit patency.

A related operational pitfall is assuming that “no alarm” means “no problem.” Some issues—like partial cannula displacement, mouth breathing that reduces perceived benefit, or early skin irritation—may not trigger device alarms but can still lead to ineffective therapy or preventable harm if not identified through bedside assessment.

What if something goes wrong?

When problems occur, a structured response reduces risk and downtime. Always prioritize patient assessment and follow local escalation policies.

Troubleshooting checklist (practical and non-brand-specific)

Patient-first checks

  • Assess the patient’s current status and tolerance.
  • Ensure the cannula is positioned correctly and not occluded (secretions, kinking, displacement).
  • Check for obvious discomfort, nasal bleeding, or skin pressure issues.

Device and circuit checks

  • Confirm the device is powered and not displaying a fault code.
  • Verify tubing connections at the device outlet, humidifier chamber, and cannula.
  • Check for circuit kinks, compression under bed rails, or pooling condensate causing partial blockage.
  • Confirm humidifier chamber water level and correct seating; replace if leaking or cracked.

Gas supply checks

  • Confirm wall outlets are active and hoses are connected to the correct ports.
  • Check oxygen and air supply pressures/indicators (as applicable).
  • If using cylinders, confirm regulator type, cylinder pressure, and expected run-time at current demand.

Alarm response

  • Read the alarm message; do not assume all alarms are minor.
  • If the alarm repeats after basic checks, consider switching to backup equipment and escalating.

In many facilities, a small number of failure patterns account for a large proportion of calls: empty water chambers, mis-seated chambers, disconnected heater wires, kinked tubing under bed rails, and low supply pressure during peak unit demand. Capturing these patterns in an internal “top issues” list can guide targeted staff refreshers and reduce repeat incidents.

When to stop use (general)

Stop use and transition per protocol when:

  • The device cannot deliver the required flow/FiO₂ due to supply failure or malfunction.
  • There is evidence of overheating, burning smell, fluid ingress into the device, or electrical safety concern.
  • The patient cannot tolerate the interface or appears to deteriorate despite appropriate monitoring and adjustments.
  • Your facility’s policy indicates a safety threshold has been reached requiring escalation to another support pathway.

When to escalate to biomedical engineering or the manufacturer

Escalate when you encounter:

  • Recurrent unexplained fault codes or alarms
  • Suspected sensor failure or calibration drift (if applicable)
  • Damaged connectors, heater wires, or internal fan/turbine noise changes
  • Frequent circuit overheating/underheating despite correct setup
  • Any event requiring incident reporting, quarantine, or recall checks

Capture device identifiers (asset tag, serial number) and retain consumable batch/lot details if your quality system requires it.

For service efficiency, it often helps to document not only the fault code, but also the settings at the time, the gas source configuration (wall oxygen/air versus cylinder), whether a new circuit was used, and whether the problem persists after swapping consumables. This information can shorten troubleshooting time and helps determine whether the issue is device-related, consumable-related, or infrastructure-related.

Infection control and cleaning of High flow nasal cannula system

Infection prevention for High flow nasal cannula system combines standard equipment hygiene with respiratory-therapy-specific controls. Always follow the IFU and your local infection prevention policy; they may be more stringent than general guidance.

Cleaning principles (what should be true everywhere)

  • Treat patient-contact components (cannula, breathing circuit, water chamber where applicable) as high-risk items and handle as directed—often single-patient use.
  • Prevent backflow and contamination from condensate by keeping tubing routed appropriately and draining away from the patient when permitted by protocol.
  • Use only approved cleaning agents and contact times; incompatible chemicals can crack plastics, fog screens, or degrade seals.
  • Separate “dirty-to-clean” workflows to prevent cross-contamination during changeover and transport.

Because high-flow therapy involves humidification, it also tends to generate more moisture-containing surfaces (circuits, chamber docks) than dry oxygen delivery. This makes consistent turnaround cleaning and careful handling of condensate especially important. Facilities often treat drained condensate as potentially contaminated fluid and avoid practices that could aerosolize it.

Disinfection vs. sterilization (general)

  • Cleaning removes visible soil and reduces bioburden; it is a prerequisite for effective disinfection.
  • Disinfection (often low-level for external surfaces) is commonly used for noncritical device surfaces.
  • High-level disinfection or sterilization is generally reserved for items that contact mucous membranes or enter sterile body sites, but with High flow nasal cannula system many of these components are disposable. Reprocessing requirements vary by manufacturer and local regulation.

Do not assume any component is reprocessable unless the manufacturer explicitly permits it and your facility has validated workflows.

High-touch points to include in routine wipe-downs

Typical high-touch points on the base unit include:

  • Touchscreen and control knobs/buttons
  • Handles, pole clamps, and cart rails
  • Power switch and power cord contact points
  • Gas inlet connectors and strain relief areas
  • Humidifier lid/locking mechanism and any water chamber docking area

Example cleaning workflow (non-brand-specific)

  1. Don appropriate PPE per facility policy.
  2. Place the device in standby/off mode and disconnect from the patient safely.
  3. Remove and discard single-use cannula and circuit items in the correct waste stream.
  4. Remove the water chamber and dispose of or reprocess it exactly as directed by the IFU.
  5. Inspect the base unit for spills, cracks, or residue; report damage.
  6. Clean exterior surfaces with facility-approved detergent/disinfectant wipes using the specified wet contact time.
  7. Pay attention to crevices, around knobs, and under handles; avoid fluid ingress into vents.
  8. Allow surfaces to air dry; do not towel-dry if it compromises disinfectant contact time.
  9. Fit new consumables when the device is next needed, or store in a clean area with ports protected.
  10. Document cleaning, consumable changeover, and any faults in the equipment log.

Water and humidifier management (often overlooked)

  • Use only the water type specified (often sterile or distilled; varies by manufacturer and policy).
  • Avoid “topping up” in a way that increases contamination risk; follow change intervals.
  • Manage condensate using safe handling steps; avoid draining toward the patient.
  • If a spill occurs inside the device housing, remove the unit from service and involve biomedical engineering.

Where respiratory infection risk is high, facilities may also add administrative controls such as limiting high-flow therapy to certain rooms, applying additional environmental cleaning frequency, or requiring specific PPE for circuit changes. These decisions are local and should be driven by infection prevention leadership in collaboration with clinical teams.

Medical Device Companies & OEMs

Manufacturer vs. OEM (Original Equipment Manufacturer)

In the medical device ecosystem, the manufacturer is the legal entity responsible for the finished product’s regulatory compliance, labeling, post-market surveillance, and overall quality management system. An OEM may supply subassemblies (for example, heaters, sensors, blowers/turbines, humidifier modules, or molded consumables) that are integrated into the final system, sometimes under private-label arrangements.

For High flow nasal cannula system procurement and lifecycle management, OEM relationships matter because they can affect:

  • Availability and lead times for spare parts and consumables
  • Service documentation quality and access to authorized repair channels
  • Software/firmware update pathways and cybersecurity responsibilities (where applicable)
  • Consistency of consumable compatibility across product revisions

When evaluating a supplier, confirm who the legal manufacturer is, what service network exists in your region, and how warranty and recalls are handled.

In addition, procurement teams often benefit from clarifying whether accessories are backward-compatible across generations of the same product line. Even small changes—such as updated connectors, revised water chambers, or new circuit SKUs—can create hidden complexity if a hospital has a mixed fleet.

Top 5 World Best Medical Device Companies / Manufacturers

The list below is example industry leaders commonly associated with respiratory care and hospital oxygen therapy portfolios. It is not a verified ranking, and product availability varies by country and regulatory approval status.

  1. Fisher & Paykel Healthcare
    Generally recognized for respiratory humidification and oxygen therapy consumables and systems used in acute care. The company is often referenced in discussions around humidified high-flow therapy, with a strong presence in many hospitals globally. Portfolio emphasis and local support coverage vary by region.

  2. Vapotherm
    Known for focusing on high-flow therapy systems and related disposables in certain markets. Often discussed in facilities that prioritize compact setups and high-flow-specific workflows. Availability, installed base, and service model vary by country.

  3. Dräger
    A long-established manufacturer in critical care, anesthesia, and ventilation. In many hospitals, Dräger’s footprint is strongest in ventilators and monitoring, and high-flow therapy may be delivered through standalone systems or via ventilator-supported modes depending on product line (varies by manufacturer configuration). Global service and training infrastructure is a common procurement consideration.

  4. Philips
    A broad medical equipment manufacturer with significant respiratory care history in many regions. Depending on the portfolio available locally, hospitals may encounter Philips in oxygen therapy accessories, monitoring ecosystems, and respiratory support devices where high-flow functionality may be offered in specific configurations. Current availability and support terms are country-specific.

  5. Hamilton Medical
    Commonly associated with ICU ventilators and respiratory support platforms. Some facilities leverage ventilator-based high-flow or oxygen therapy modes as part of broader respiratory workflows (capabilities vary by model and regulatory clearance). Procurement teams often evaluate Hamilton within a “platform strategy” that includes serviceability and clinician training.

For procurement teams comparing manufacturers, it can be useful to look beyond headline specifications and ask practical lifecycle questions such as: How many consumable SKUs will we stock? Are there local service engineers? What is the typical turnaround time for repairs? Is there a loaner program? Are alarms and user interface consistent across models? These factors often drive real-world safety and continuity more than the published maximum flow value.

Vendors, Suppliers, and Distributors

Role differences between vendor, supplier, and distributor

In healthcare procurement, these terms are often used interchangeably, but they can describe different roles:

  • Vendor: The entity that sells to the hospital (may be the manufacturer, an authorized reseller, or a tender-awarded partner).
  • Supplier: A broader term for any organization providing goods/services, including consumables, spare parts, and service labor.
  • Distributor: Typically holds inventory, manages logistics, and may provide first-line technical support, returns processing, and coordination with manufacturers for warranty service.

For High flow nasal cannula system, understanding who is authorized to supply consumables (cannulas, circuits, water chambers) is as important as who supplies the base unit. Non-authorized consumables can introduce fit, performance, and infection control risks, and may affect warranty terms (varies by manufacturer).

In addition to authorization status, hospitals often evaluate distributors on their ability to support continuity of supply during demand spikes. For example, having local buffer stock of circuits and chambers, clear lead times, and a transparent substitution policy (when substitutions are even allowed) can prevent last-minute clinical workarounds.

Top 5 World Best Vendors / Suppliers / Distributors

The list below is example global distributors often referenced in broader hospital supply chains. It is not a verified ranking, and whether they supply High flow nasal cannula system specifically depends on country, tendering structures, and authorized distribution agreements.

  1. McKesson
    A major healthcare distribution organization in certain markets, often supporting hospitals with medical-surgical logistics, inventory programs, and procurement services. Product availability for respiratory therapy equipment varies by region and contracting. Typically serves large provider networks and integrated delivery systems.

  2. Cardinal Health
    Commonly involved in hospital supply distribution and value-added logistics in selected countries. May support respiratory care categories alongside broader consumables and supply chain services. Specific high-flow product portfolios depend on local authorization and contracts.

  3. Medline
    Widely known for medical-surgical distribution and a large consumables portfolio in multiple regions. Often relevant to procurement teams building standardization programs for disposables and infection prevention supplies that interact with respiratory workflows. Availability of specific branded high-flow systems varies.

  4. Owens & Minor
    Often associated with healthcare logistics, inventory management, and distribution services in certain markets. Can be relevant where hospitals outsource portions of supply chain operations and require consistent consumable replenishment. Product access and service models are contract- and country-dependent.

  5. DKSH
    A distribution and market-expansion service provider with a presence in parts of Asia and other regions. Often works as an in-country partner for medical device manufacturers, supporting regulatory, sales, and after-sales functions. Coverage and device categories vary by country and manufacturer agreements.

Global Market Snapshot by Country

India

Demand for High flow nasal cannula system is influenced by expanding ICU capacity, high respiratory disease burden, and ongoing investments in oxygen infrastructure after recent public health surges. Imports remain important for many base units and branded consumables, while service coverage is stronger in major urban hospitals than in smaller districts. Public and private procurement pathways can differ significantly, and hospitals often prioritize devices with strong local training and spare parts availability.

China

China combines high volume demand with an extensive domestic medical equipment manufacturing ecosystem, leading to both imported and locally produced high-flow options. Large tertiary hospitals in urban centers typically have stronger procurement leverage and service coverage, while rural access can be constrained by infrastructure, staffing, and standardization gaps. Domestic supply capacity can support scale, but product selection is still shaped by regulatory pathways and local hospital tender requirements.

United States

In the United States, High flow nasal cannula system adoption is supported by mature critical care pathways, strong distributor networks, and established service expectations. Procurement decisions often emphasize total cost of ownership, consumable contracts, training, and integration with respiratory therapy workflows across ED, ICU, and step-down units. Facilities may also consider how high-flow programs align with staffing models (respiratory therapists vs nurse-led protocols) and alarm management in different care areas.

Indonesia

Indonesia’s demand is concentrated in urban referral hospitals, with growing interest in noninvasive oxygen therapy that can be deployed outside ICUs when monitoring allows. Import dependence is common for premium systems and consumables, and service capability can vary significantly across the archipelago. Logistics planning is particularly important for consumables because inter-island lead times can be unpredictable.

Pakistan

Pakistan’s market is shaped by variable oxygen infrastructure, differences between private tertiary centers and public hospitals, and the practical need for robust devices that tolerate challenging environments. Many facilities rely on imported systems, with procurement often balancing upfront cost, consumable availability, and local technical support. Oxygen supply planning and cylinder logistics can strongly influence deployment beyond major centers.

Nigeria

In Nigeria, access is uneven, with higher uptake in private and federal tertiary facilities than in rural settings. Oxygen supply reliability and biomedical engineering capacity are key constraints, making serviceability, spare parts access, and training central to procurement of High flow nasal cannula system. Programs that pair device procurement with oxygen infrastructure support and competency training tend to be more sustainable.

Brazil

Brazil has strong clinical demand in large urban hospitals and an established medical device distribution ecosystem, though procurement pathways can differ across public and private sectors. Importation plays a significant role for some brands, and service support quality may vary by region and distributor capability. Standardization across hospital networks can be a key driver of consumable pricing and continuity.

Bangladesh

Bangladesh’s demand is driven by dense urban patient volumes and growing critical care services, with significant reliance on imported devices and consumables. Facility constraints can include oxygen supply capacity, staffing ratios, and the need for standardized training to safely expand high-flow therapy beyond ICUs. Distributor reliability and consumable lead time management are often decisive in product selection.

Russia

Russia’s market reflects a mix of domestic production and imports, with procurement influenced by regulatory requirements, supply chain constraints, and regional variability in hospital modernization. Large metropolitan centers typically have stronger service ecosystems than remote regions, affecting deployment consistency for High flow nasal cannula system. Procurement may also emphasize long-term parts availability due to geographic and logistics considerations.

Mexico

In Mexico, adoption is stronger in urban hospitals and private networks with established respiratory therapy services. Import dependence is common for many high-flow platforms, and procurement teams often focus on distributor support quality, consumable continuity, and maintenance turnaround times. Hospitals may evaluate whether devices can be supported effectively outside major cities where service engineers are concentrated.

Ethiopia

Ethiopia’s demand is increasing with critical care development and broader investments in oxygen systems, but access remains concentrated in major cities. Import reliance and limited biomedical engineering bandwidth can make training, spare parts planning, and simple maintainability especially important for High flow nasal cannula system programs. Facilities often benefit from devices that are resilient to power and infrastructure variability.

Japan

Japan’s market is characterized by high standards for hospital equipment reliability, structured clinical governance, and strong expectations for vendor service and training. Adoption is supported by advanced acute care systems, though purchasing decisions may be influenced by domestic distribution structures and strict quality requirements. Lifecycle documentation and service response commitments often weigh heavily in procurement decisions.

Philippines

In the Philippines, demand is centered in Metro Manila and other major cities, with variability across island regions due to logistics and infrastructure. Importation is common, and facilities often prioritize dependable service partners, consumable supply continuity, and oxygen capacity planning when scaling High flow nasal cannula system use. Regional distribution capabilities can strongly affect standardization across multi-hospital networks.

Egypt

Egypt shows growing demand in large public and private hospitals, with procurement often balancing cost controls against the need for reliable oxygen therapy equipment. Import dependence is common for many branded systems, and service ecosystem strength can differ between major urban centers and peripheral areas. Hospitals may place added emphasis on consumable pricing stability and training support.

Democratic Republic of the Congo

In the Democratic Republic of the Congo, access to High flow nasal cannula system is limited by infrastructure constraints, supply chain complexity, and uneven distribution of trained staff. Where deployed, sustainability depends heavily on oxygen reliability, consumable availability, and support from partners with local technical presence. Programs may need to account for longer lead times and limited repair capacity.

Vietnam

Vietnam’s market is expanding with continued hospital modernization and increased critical care capability in major cities. Imports remain important for many high-flow systems, while local distribution networks are improving; rural access and consistent biomedical service coverage remain key differentiators. Hospitals often evaluate how quickly consumables can be replenished outside the largest urban hubs.

Iran

Iran has significant clinical demand and a mix of local production capability and imports, shaped by regulatory and supply chain realities. Hospitals may prioritize systems with strong local serviceability and stable consumable supply, especially where international procurement timelines are uncertain. Standardization may be influenced by local manufacturing partnerships and in-country technical support.

Turkey

Turkey’s demand is supported by a large hospital sector and a growing medical device distribution and service ecosystem. Procurement often emphasizes robust after-sales support and reliable consumables, with strong adoption in urban centers and variable access in more remote regions. Competitive tendering can place additional focus on warranty terms and training coverage.

Germany

Germany’s market reflects strong clinical governance, established procurement processes, and high expectations for device safety, documentation, and service. Adoption of High flow nasal cannula system is typically supported by mature oxygen infrastructure and trained respiratory/critical care teams across many hospitals. Procurement teams often require detailed technical documentation and clear preventive maintenance pathways.

Thailand

Thailand’s demand is concentrated in Bangkok and major provincial centers, with expanding critical care services and attention to noninvasive oxygen therapy options. Import dependence is common for many systems; hospitals often evaluate local distributor service depth, training quality, and consumable continuity for sustainable rollout. As use expands, oxygen infrastructure assessment can become a gating factor for multi-bed implementation.

Key Takeaways and Practical Checklist for High flow nasal cannula system

  • Treat High flow nasal cannula system as both oxygen delivery and humidification hospital equipment.
  • Confirm the legal manufacturer and model-specific IFU before standardizing any workflow.
  • Verify oxygen infrastructure capacity before scaling high-flow therapy across multiple beds.
  • Plan for peak oxygen demand scenarios, not average day-to-day consumption.
  • Stock consumables (cannulas, circuits, chambers) based on realistic burn rates and lead times.
  • Standardize cannula sizing guidance and stocking across units to reduce setup errors.
  • Build competency training that covers assembly, alarms, humidification, and oxygen safety.
  • Require documented sign-off for new staff before independent operation.
  • Include biomedical engineering in device selection to assess serviceability and PM requirements.
  • Confirm whether FiO₂ is measured or only set; interpretation and verification differ.
  • Use a structured pre-use check to catch missing water chambers and misconnected hoses.
  • Label gas hoses and wall outlets clearly to reduce misconnections and downtime.
  • Keep circuits routed to minimize pooling condensate and accidental occlusion.
  • Treat condensation management as a safety task, not just a comfort issue.
  • Ensure alarms are audible in the intended care environment and not routinely silenced.
  • Investigate recurring nuisance alarms as a training or compatibility problem.
  • Document initial settings, setting changes, and alarm events per local policy.
  • Confirm the device has passed electrical safety checks within your facility schedule.
  • Avoid unapproved third-party consumables unless formally validated by the facility.
  • Protect skin at cannula contact points and check pressure areas at routine intervals.
  • Ensure the humidifier is filled with the specified water type and to the correct level.
  • Do not “top up” water in ways that conflict with infection prevention policy.
  • Use only approved disinfectants on screens and plastics to prevent material damage.
  • Clean high-touch surfaces (screen, knobs, handles) every turnaround as a minimum.
  • Replace damaged tubing, connectors, and heater wires rather than improvising repairs.
  • Keep a backup oxygen delivery method available for any bed using high-flow therapy.
  • For transport, calculate oxygen supply needs and confirm battery duration in advance.
  • Secure cylinders, carts, and tubing to prevent falls, disconnections, and regulator damage.
  • Maintain a clear escalation pathway when patient response is inadequate or uncertain.
  • Separate patient physiology monitoring from device status; the device does not assess the patient.
  • Track device utilization and failure modes to inform preventive maintenance planning.
  • Maintain a log of fault codes and repairs to identify repeat issues and training gaps.
  • Confirm local availability of authorized service and spare parts before purchase.
  • Include service response times and loaner policies in contracts where possible.
  • Align procurement with infection control requirements for disposable vs reusable components.
  • Use incident reporting for overheating, fluid ingress, repeated faults, or suspected device failure.
  • Quarantine devices with unresolved faults and involve biomedical engineering promptly.
  • Review room ventilation and PPE policies where respiratory infection risk is high.
  • Treat oxygen enrichment as a fire risk and enforce ignition source controls consistently.
  • Reassess ward-level deployment carefully; monitoring and staffing models must match risk.
  • Validate that any software/firmware updates follow your facility’s change-control process.
  • Harmonize device fleets where practical to reduce training burden and consumable variation.
  • Plan end-of-life replacement and parts obsolescence to avoid sudden program disruption.
  • Map which bed spaces have compatible oxygen/air outlets (and outlet flow capability) before expanding service lines.
  • Use simple oxygen demand estimates during planning to avoid unexpected pipeline strain when multiple beds run high settings simultaneously.
  • Standardize who changes water chambers and how often, and audit compliance during busy periods.
  • Create a quick-reference guide for your specific device alarms and the first-response steps expected on your unit.
  • Maintain a small “ready kit” (spare chamber, circuit, cannula, wipes) to speed turnaround and reduce therapy downtime.